Appendix B: Performance Verification
BĆ22
DG2030 User Manual
The following steps check the external trigger signal.
40. Press the following buttons to set the DG2030 trigger source to Ext.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Trigger
Source
(Set to Ext)
41. Change the connection as follows.
a. Disconnect the BNC cable connecting between the DG2030 rear panel
EVENT OUTPUT connector and the oscilloscope CH3 input connector.
b. Disconnect the BNC cable connecting between the DG2030 rear panel
EVENT INPUT connector and the function generator output connector.
c.
Connect a BNC T adaptor to the function generator output connector.
And using two BNC cables, connect from the BNC T adaptor to the
DG2030 front panel TRIGGER INPUT connector, other side BNC T
adaptor to the oscilloscope CH3 input connector.
d. Disconnect the BNC cable from the DG2030 front panel CH3 connector
and re–connect it to CH2 connector. (Disconnect the BNC cable from the
DG2030 front panel CH7 connector and re–connect it to CH2 connector
if the option 01 has been installed.)
e.
Disconnect the BNC cable from the DG2030 front panel CLOCK OUT
connector and re–connect it to CH0 connector.
See Figure B-9.