Appendix A: Specifications
AĆ2
DG2030 User Manual
Table AĆ1: Warranted electrical characteristics
Characteristics
Description
Performance test
Internal trigger generator
Internal trigger rate
Accuracy
±
0.01 %
Data and clock out
*Output voltage
See page BĆ28.
Accuracy
DC (data out)
(
±
3 % of setting)
±
50 mV (into 50
W
)
Amplitude (clock out)
(
±
5 % of setting)
±
50 mV (into 50
W
)
Maximum Output Current
±
100 mA
*Rise/fall time (20 % to 80 %)
See page BĆ28.
Accuracy
(
±
10 % of setting)
±
500 ps, within the variable range
*Channel skew
t
±
300 ps, reference to the clock out, at 10 MHz, 50 % duty
and rise/fall times set to FAST.
See page BĆ33.
*Delay function
See page BĆ33.
Accuracy
(
±
3 % of setting)
±
500 ps
±
6
0 ps
| Ambient temperature (
_
C)
*
25 |,
At 10 MHz, 50 % duty, 1.5 V high, 0 V low, and rise/fall times set
to FAST.
Event input
Minimum pulse width
y
100 ns
Sensitivity
y
1.0 VpĆp
Inhibit input
Minimum pulse width
y
100 ns
Sensitivity
y
1.0 VpĆp
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com