Appendix B: Performance Verification
BĆ24
DG2030 User Manual
44. Set the oscilloscope settings to:
Horizontal axis
400 ns/div
. . .
Record length
5000
. . . .
Vertical axis
1 V/div (for CH3 input)
. . . . .
Acquisition mode
Run/Stop button only (Stop After menu)
.
Input impedance
50
W
(for CH3 input)
. .
Trigger level
500 mV
. . . . .
Trigger source
CH4
. . . .
45. Observe the displayed waveform patterns on the oscilloscope screen, confirm
that the the trigger event synchronizes with the rising edge of the input pulse.
Trigger event can be recognized by observing the CH1 output signal from
the DG2030 on oscilloscope CH1 input.
46. Press the following buttons to set the DG2030 trigger slope to Negative.
Menu button
Bottom button
Pop up menu
Side button
Front panel
button
SETUP
Trigger
Slope
(Set to Negative)
47. On the oscilloscope screen, confirm that the trigger event occurs synchroniz-
ing with the falling edge of the input pulse.
48. Press the START/STOP button on the DG2030 front panel to stop output
(the button LED goes off).