Appendix B: Performance Verification
DG2030 User Manual
BĆ35
4. Set the delay time to –5.00 ns, 0.00 ns, +5.00 ns, +10.00 ns, +15.00 ns and
+18.00 ns in turn. Perform the delay time measurement for each delay time
setting with the oscilloscope. Confirm that the data output delay times
referenced to the clock signal are within the specified range for each of these
delay time settings.
5. Change the BNC cable connection from the CH0 to CH1 through CH3 for
standard (or CH7 for option 01) in turn and repeat step 4 for each of these
channels.
6. From above measurement results at 0.00 ns delay time setting on all output
channels, confirm that the delay time differences between two adjacent
channels is within the range from –300 ps to +300 ps.
7. Press the START/STOP button on the front panel to stop data output (the
button’s LED goes off).