Utility Menu
DG2030 User Manual
3Ć89
Operation.
Set the directory listing order
Bottom button
Popup menu
Side button
Mass Memory
Special
Catalog Order (Select
the display type)
Operation.
Lock a file
Bottom button
Popup menu
Side button
Mass Memory
Special
Lock (Select On)
Cal Menu
Performs the calibration for the internal clock circuit. When the calibration
finishes without error, “Pass” is displayed in the Status display area. If an error
occurs, “Fail” is displayed and an error code appears in the Comment display
area.
Operation.
Lock a file
Bottom button
Popup menu
Side button
Cal
Execute
Table 3Ć21: Error Code
Error code
Error meaning
Failed component
8XX
CAL error
Clock board
The clock calibration is recommended to be performed in the following cases:
H
After a 20-minute warm up period has been passed.
H
When you need data pattern output with high precision frequency.
H
When the ambient temperature is changed by about
±
5
_
C since the previous
clock calibration.