<DelayTime> Pulse delay time of pulse modulation.
Range: Non-Trig Mode: 0s [0ns
,
42.000000000s],
Trig Style: 0s [100ns
,
42.000000000s].
Programmable command:
[:SOURce]:PULM:INTernal:DELay <val>
[:SOURce]:PULM:INTernal:DELay?
Example:
:PULM:INTernal:DELay 1ms Set pulse delay to 1 ms
4.2.4.6 PRF
Functional Description:
This command sets PRF of pulse modulation. When Source is Square, the pulse signal outputs
a signal with duty cycle of 50%, and the command can change the frequency of square signal.
Parameter specification:
<Frequency> PRF of pulse modulation.
Range: 1kHz [0.023Hz
,
25MHz]
Programmable command:
[:SOURce]:PULM:INTernal:FREQuency <val>
[:SOURce]:PULM:INTernal:FREQuency?
Example:
:PULM:INTernal:FREQuency 1MHz
Set PRF to 1 MHz.
4.2.4.7 Input Direction [ON/OFF]
Functional Description:
The command logically inverts the external input pulse signal, i.e. when Pulse Source is EXT,
the pulse signal input from pulse input port on the front panel of the signal generator is TTL high level
signal or is inverted to TTL low level signal.
Parameter specification:
<Mode> Discrete data. Input Direction is as follows:
NORMal Input Direction is turned off and the input pulse signal is TTL highlevel.
INVerted Input Direction is turned on and the input pulse signal is TTL low level.
Programmable command:
[:SOURce]:PULM:EXTernal:POLarity INVerted|NORMal
[:SOURce]:PULM:EXTernal:POLarity?
Example:
:PULM:INTernal:FREQuency 1MHz
Set PRF to 1 MHz.
4.2.4.8 Delayed Sweep
Functional Description:
Not implemented
4.2.4.9 AM [ON/OFF]