Tel: 886.909 602 109 Email: [email protected]
148
4.2.5.5 Filter Select
Functional Description:
This command selects the baseband pre-modulation filter types of the signal generator,
including: RNYQuist, NYQuist, GAUSsian, RECTangle.
Parameter specification:
<Mode>
Discrete data. Baseband pre-modulation filter types are as follows:
RNYQuist
Root Nyquist filter,
NYQuist
Nyquist filter,
GAUSsian
Gauss filter,
RECTangle
Rectangle filter.
Programmable command:
[:SOURce]:RADio:CUSTom:FILTer RNYQuist|NYQuist |GAUSsian| RECTangle
[:SOURce]:RADio:CUSTom:FILTer?
Example:
:RADio:CUSTom:FILTer RNYQuist
Baseband pre-modulation filter type is Root Nyquist.
4.2.5.6 Optimize Filter
Functional Description:
Not supported.
4.2.5.7 Trig Style
Series S1465 Signal Generator has three trigger modes for real-time baseband: Continue,
Single and Gate. The function of each style is described as follows:
Continue: After a valid trigger event is received, the currently selected code pattern is repeatedly
generated.
Single: After a valid trigger event is received, the currently selected code pattern is generated once.
Gate: The current code pattern is generated continuously and repeatedly during the effective
period of gate signal.
Functional Description:
This command sets the baseband signal trigger mode that controls data transmission, including:
CONTinuous, SINGle and GATE.
Parameter specification:
<Mode>
Discrete data. Baseband signal trigger style is as follows:
CONTinuous
| 0: Trig Style is set to Continue,
SINGle
| 1: Trig Style is set to Single,
GATE
| 2: Trig Style is set to Gate.
Programmable command:
[:SOURce]:RADio:CUSTom:TRIGger:TYPE CONTinuous|SINGle|GATE
[:SOURce]:RADio:CUSTom:TRIGger:TYPE?
Example:
[:SOURce]:RADio:CUSTom:TRIGger:TYPE SING
Trig Style is Single.
4.2.5.8 Trig Source
Trig source of real-time baseband of 1465 series signal generator includes: Key, Ext and Bus.
Each trigger source is described as follows.