Joint surface
N type(female)
N type (male)
Fig.2.8 Calibration Surface
2.1.1.5 User checks
After the initial power-up of the S1465 series signal generator, it is necessary to check whether
the instrument is functioning properly in order to continue measurements.
NOTE
Instructions on using the hard keys on the front panel and the soft menu keys.
1)
Hard key:
【
XXX
】
, XXX denotes the name of the hard key.
2)
Soft key:
Soft key: [XXX], XXX denotes the name of the soft key
1)
Self-test
Power on the S1465 series signal generator, and check that the power indicator lamp above the
power button at the bottom left corner of front panel turns yellow, now the standby power supply is
working properly. Press the power button on the front panel, observe that the power indicator lamp on
the front panel turns green and the display backlight lights up, and wait for about 30 s to display the
power-on status interface.
After warming up the signal generator for 10min, set it by following the steps below:
Step 1. Press the
【
System
】
key on the front panel or click the [System] function area on
the interface to enter the system menu;
Step 2. Select [Instrument Self Test] in the self test option;
Step 3. In the pop-up Self Test Config Dialog, select self test items and click [Start Test],
then observe the test result:
If the test is successful, a message “All pass self test”
will be displayed, indicating the instrument is working properly; if not, a message “Self test failed in
XX” will be displayed, indicating the instrument is not working properly. In this case, contact our
service consultation center via the contact provided on front cover II or in “6.4 Repair
method” for quick repair or replacement depending on the situation.
2)
Functional verification
Start and warm up the instrument for at least 30min, and apply a matched load to RF output. Set the
instrument as below:
Step 1. Press the
【
Amplitude
】
key on the front panel or click the [Amplitude] function area
on the interface to enter the Power menu and set the power to 0dBm in the pop-up Power
Parameter Setting Dialog.
Step 2. Press the
【
Frequency
】
key on the front panel or click the [Frequency] function area on
the interface to enter the Frequency menu and set CW to 100MHz in the pop-up Frequency
Parameter Setting Dialog.
Step 3. Press the
【
RF ON/OFF
】
key on the front panel or click the [RF] function area on the
interface to turn on the RF output;
Step 4. Press the arrow keys on the front panel to set the frequency in increments of 100MHz
until the maximum value is reached. Then observe whether there is any alarm indication in the