<Ampl>
LF output signal amplitude.
Range: 2.000VPP[0.002VPP
,
4.000VPP].
Programmable command:
[:SOURce]:LFOutput:AMPLitude <val>(unit:VPP|Mvpp|VRMS)
[:SOURce]:LFOutput:AMPLitude?
Example:
:LFOutput:AMPLitude 1VPP
LF output signal amplitude is set to 1 VPP.
4) LF DC Offset
Functional Description:
Not supported
5) LF Waveform
Functional Description:
This command sets LF signal output waveform. There are seven waveforms for selection: Sinc,
Square, Triangle, Zigzag, White noise, SweepSinc and DualSinc.Select this menu, and the menu
item will be lighted, and the waveform of LF output signal is Sinc.
Parameter specification:
<Mode> Discrete data. LF Waveform is as follows: SINE
Sine,
SQUare
Square,
TRIangle
Triangle,
RAMP
Zigzag,
NOISe
Noise,
SWEPtsine
SweepSinc,
DUALsine
DualSinc.
Programmable command:
[:SOURce]:LFOutput:SHAPe SINE|SQUare|TRIangle|RAMP
|NOISe|SWEPtsine|DUALsine [:SOURce]:LFOutput:SHAPe
?
Example:
:LFOutput:SHAPe TRIangle
4.2.1.8 Phase Ref Config
Functional Description:
Not supported
4.2.1.9 Phase Adj
Functional Description:
Not supported
4.2.2 Ampl
4.2.2.1 Ampl
Functional Description:
This command sets the output amplitude level of the signal generator.
Parameter specification:
<Ampl> Amplitude level value.
Range: -135dBm [-135dBm
,
+30dBm].
Programmable command: