Selections are made by an ON-OFF switch in the screen and by numerical inputs.
If virtual keyboard is switched ON a virtual keyboard pops up for input the data. When Virtual keyboard is
switched OFF the arrow keys must be used for input of data.
When ready with press
Save settings
before you continue.
The following selections can be switched ON or OFF:
1.
30 min. method
▪
30 min. method
ON: ESR's are measured after 30 minutes.
▪
30 min. method
OFF: ESR's are measured after 60 minutes.
2.
EDTA mode
▪
EDTA mode
ON: Samples are presented in EDTA sample tubes. The samples are diluted in the
Starrsed RS.
▪
EDTA mode
OFF: Samples are presented in pre-diluted CITRATE sample tubes. Dilution on the
Starrsed RS is switched OFF.
3.
Temp. Correction
▪
Temp. Correction
ON: A temperature corrected value for the ESR is shown besides the actual
measured value.
▪
Temp. Correction
OFF: Only the actual measured ESR value is shown.
4.
Fast filling
▪
Fast filling
OFF (Default setting): The carousel is filled with optimum pipette usage but completion of
a sample rack takes a little longer. The rotation sequence of the carousel is evenly divided. In 60
minute method, the carousel moves one position every 40 seconds (approx.). In 30 minute method,
the carousel moves one position every 20 seconds (approx.).
Fast filling
ON: The carousel is filled with
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
79
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...