Sensor check
Vacuum pressure check
•
Go to tab
Maintenance
->
Check sensor
. Select
Check Flow sensor
box.
Flow:
0980
± 60 Abs:
0320
±10
If the flow is not in range there might be a blockage in the vacuum flow line to the flow sensor.
Fill Stop sensor check
•
Go to tab
Maintenance
->
Check sensor
. Select
Check Fill stop sensor
box.
Fill stop sensor FS 90..
140
..165
Diluter Start sensor check
•
Go to tab
Maintenance
->
Check sensor
. Select
Diluter start sensor
box.
Diluter start sensor 400-
550
-700
Measure sensor check
•
Go to tab
Maintenance
->
Check sensor.
Select
Check measure sensor
box.
Measure sensor MS
50
±10
Temperature sensor check
•
Go to tab
Maintenance
->
Check sensor.
Select
Check Temperature sensor
box.
Temperature sensor TS [Room temperature]
Diluent flow sensor check
•
Go to tab
Maintenance
->
Check sensor.
Select
Check Diluent flow sensor
box.
Press test. When test is finished, signal Down and signal Up must be green.
Separator check
•
Go to tab
Maintenance
->
Check sensor.
Select
Check Separator sensor
box.
Separator sensor <200
600
>700
Clean waste separator and exchange Waste cassette assembly
Be careful, as there may be blood in the cassette. First, make up some disinfectant and put this in the
waste separator. Press
Prime Disinfectant
to pump disinfectant through the pump cassette.
Work instruction Starrsed RS
MRN-071-2-EN version 7.02 Starrsed RS
245
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...