1.
Prime
Pipettes
, activates the Rinse pump.
Rinse solution
must flow through the pipette.
2.
Prime
Needle/nozzle
, activates the Saline pump. Liquid must flush through the needle assembly.
3.
Prime
Diluter
, activates the diluter prime cycle. Diluter system must be filled with diluent and free of air
bubbles.
Diluter prime cycle is executed once. In order to fully prime the system it will be necessary to perform this
step several times. (One cycle is 5 strokes of the Diluter)
4.
Prime
Nozzle with water
, activates the fill nozzle water valve. De-ionized water must flow through the tube
connected to the fill nozzle cap.
5.
Prime
Waste system
, activates the disinfectant valve. Disinfectant must flow through the small tube
connected to the pipette wash station.
When the Starrsed RS has been idle for more than eight hours, there might be some backflow of reagents
due to gravity. Prime all tubing before sampling by using the function:
•
Prime all units
All priming functions are sequentially performed one time.
8.1.4 - Fill procedure
Fill the sample rack with the sample tubes. Observe that the barcodes are visible through the opening in the
rack. Place the rack in a rack adapter.
Set the Rack adapter in the rack loader with the barcodes facing the rack grabber. Select the tab
Sample
and
press the button Sample mode.
The Rack is transported to the rack grabber, the rack grabber picks up the Rack adapter and moves to the
barcode reader position to scan the first barcode label. The rack is then mixed and after that moved to the
needle position to be aspirated.
If the barcode labels fail to read check if the barcode is facing the correct direction.
8.1.4.1 - Liquid levels
Liquid containers and levels must be checked frequently.
If the small onboard bottles are used, wash and keep the bottles clean to avoid bacterial growth.
The Starrsed RS has liquid level sensors. When the level sensor alarm appears, replace reagent as soon as
possible.
Operation
MRN-071-2-EN version 7.02 Starrsed RS
113
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...