Make adjustment if necessary in
[Settings] -> [Diluter settings] -> [Dilution adjustment]
.If applicable:
hier hoort
ook 21490 in
4. Check your on-board diluent container Recommendation is to clean the bottle when you replace the 5 liter
container Diluent.
10. Check and clean instrument
1. Check system for leakage.
▪ Inspect the peristaltic pump tubes and connections for leaks.
▪ Check that liquid does not run back into the supply containers after the pumps have stopped.
2. Check tubing from the syringe for trapped air bubbles.
3. Check Diluent syringe for trapped air bubbles.
4. If trapped air bubbles are found, go to tab
[Maintenance],
click button
[Prime / Clean]
and perform the
[Prime
Diluent/Diluter]
function.
5. Wipe outer surface and stainless steel plate below the pipettes with disinfectant.
11. Exchange air filter
Air filter replacement QWLV040003
1. Pull both tube connectors out of the blue disk filter.
2. Place new blue disk filter
3. Reconnect the tube connectors on the filter
Work instruction Starrsed RS
248
MRN-071-2-EN version 7.02 Starrsed RS