6.7.4.3 - Valve control
•
Load settings
is used for reloading the stored software settings.
•
Save settings as
is used for storing software settings to a file. There are no restrictions for the file
name.
•
Save settings
is used for storing software settings after settings are changed or altered.
Valves
1.
Valve outer needle:
Energizes the outer needle solenoid valve.
2.
Sample control solenoid
: Sample control solenoid fill sequence energized. The function of this solenoid is
to build up a vacuum in the Westergren pipette before the aspiration cycle starts.
3.
V-Valve wash section
: Pipette wash vacuum control valve, controls the main vacuum line between the
wash-station and separator.
4.
V-Valve fill section
: Vacuum control fill-nozzle / sample probe, controls the main vacuum line between the
fill nozzle cap and separator.
Actuators
1.
Rinse actuator
: Rinse solenoid active, rinse valve-block down.
2.
Fill actuator
: Fill solenoid active, fill valve-block down.
6.7.4.4 - Valves and Actuators
Valves
Starrsed program
106
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...