ESR Error code messages:
0
No errors
1
No cells/plasma found
Error
No contents could be detected in the pipette.
2
ESR Probably > 140 mm
Error
Extremely high ESR value.
3
Too many borders found
Error
More than three borders found, possibly air
bubbles. See Section Troubleshooting
.
4
Column height <nnn>
Warning
Column height must be between 180 and
210mm. <nnn> = the actual column height.
5
Measure error
Warning
The down count is not equal to the up count
from the measure head.
6
Bubbles on top
Warning
Air bubbles on top of the ESR. See Section
7
Limit error
Error
One of the following limits are out of the setting
range:
•
ESR Time
•
Column height
•
Dilution
•
Bubbles on top
•
Hazy aspect
•
Temperature
9.2 - Result Printout
The results of the ESR measurements can be send to the printer. The layout of the report depends on the
selection of the 60- or 30 minute method.
Columns:
1. Sample ID=Patient number.
2. Hh=Not corrected 30-minute
result (only in use if 30 minute mode is active).
3. ESR=Not corrected 60-minute
result.
4. Tc=60-minute ESR result in millimeters, corrected for
18
°C. (only in use if temperature correction is
active).
5. Aspect (clear, hazy (see page 254)).
6. Pip.=Sedimentation pipette number (number on the pipette belt).
Reporting
MRN-071-2-EN version 7.02 Starrsed RS
117
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...