16.7 - Error list Starrsed RS
Last updated: 01-09-2014
Error
Extra explanation
Reason/Solution
E2: Communication error! (Board:
%s (%x), Command: %x, TWSR:
%x E: %d)
Communication lost after 3 retries
between Computer and Starrsed
RS.
•
Power cable not connected
on the communication PCB
mounted on the back panel.
•
An I2C cable not connected
•
Serial cable not connected
•
No power on one of the
PCB's
•
Short circuit or fault on one
of the PCB's
E3: Measure motor timeout!
Measure head motor did not move
or motor is blocked.
•
Measure head is not at the
Home position.
•
Check the Home sensor.
•
Motor is faulty.
•
Motor driver on drive board
is faulty.
E4: Sample probe not in top
position! (home)
Sample probe not back at Home
position after sampling a tube.
•
Check sample probe home
sensor.
•
Sample probe motor is
faulty.
•
Sample probe motor driver
on needle board is faulty.
•
Sample probe is blocked.
E5: Duplicated ID !!
Sample rejected. Sample already in
carousel.
•
Wait until sample is
measured
•
Check general settings
(Check for duplicate ID's)
E6: Program was not properly shut
down. Check settings before
continuing!
There is a possibility that changed
settings which were not saved to
disk are lost.
•
Program stopped and
computer needed to be reset.
•
Computer reset after power
failure.
Appendix for Starrsed RS
186
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...