6.3.5.7 - QC Result analysis
Authorized staff should identify and differentiate acceptable/unacceptable random errors and trends and/or
shifts in systematic errors from the statistical data. Depending on the users Quality Control Procedures
analytical results could be accepted or rejected.
Changes in QC results can be gradual or abrupt. Gradual changes can be caused by contamination and
incidental environmental variations. Abrupt changes can be caused by change of QC material batch or
possible hardware errors.
If results are continuously out of range due to significant difference between calculated mean and control
value, but the statistics show precise results with small deviations, it should be considered to expand the
acceptable assay range with
.
If results are incidentally out of range it is advised to perform a daily maintenance and/or fill and clean step
and then perform another QC sample step before releasing patient results.
If results are not sent to the LIMS QC Results can be exported to MS Excel CSV files for further analysis in
lab's own Quality Control data system.
6.3.6 - History sample analyse option
Make a selection for all or today's results between start time and end time or make a selection for:
1. A specific number of past days.
2. A specific date.
3. A range of days from start date to end date.
Starrsed program
58
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...