6.6.5 - QC Settings
In "QC settings" the following options can be selected:
Use default assay range
ON (=Default setting): The default assay range which is predetermined for the current batch of Starrsed
Control is used. These ranges cannot be changed in this option.
OFF:
Use custom assay range
.
Use custom assay range
ON: The lab can establish their own acceptable ranges, both ranges can be set from a minimum of +/- 2
mm to a maximum of +/- 15 mm.
OFF:
Use default assay range
.
Note: It is advised to use the default assay ranges. Use caution when setting the custom ranges. A too
narrow range may cause unjustified rejection of QC sample results, subsequent rejection of patient
results and undue burden on maintenance.
Send QC request to LIMS
▪
ON: A sample request for the QC sample is send to LIMS. The QC sample will only be processed if the
LIMS responds with YES.
Use this setting if the ESR result of the QC sample will be send to LIMS as well and LIMS requires that all
samples are requested first.
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
91