1. Check the room temperature near the pipettes with a room thermometer.
2. Enter the measured temperature from the room thermometer as the current value for the temperature
sensor.
3.
"Save settings"
4.
Tip:
Measure the temperature, preferable when the instrument is not used for a longer period.
Recommended is to measure the current room temperature by holding the thermometer in the carousel and
before starting normal operation.
15.26 - Communication error (E2)
Error E2 appears if there is a communication problem between the Starrsed RS and computer. During startup
3 retries are performed, the software checks the availability of all needed connections.
Example of the message:
E2: Communication error! (Board: Needle (56), Command: 28, TWSR: 0, E: 3)
The indication of the board gives the first location to check, but because all PCB's are internally connected
the error can be originated on another location in the instrument.
The other codes (command: XX, TSWR: XX and E:XX) are indicators for status/actions from software and
I2C connections. Each board has its own range of commands and can not be explained in one way.
Most common solutions:
•
Power cable not connected on the communication PCB mounted on the back panel.
•
An I2C cable not connected
•
Serial cable not connected
•
No power on one of the PCB's
•
Short circuit or fault on one of the PCB's
15.27 - Quality control troubleshooting
Error messages
E115: QC expired, not sampled!
The used Starrsed Control is out of date, no ESR result is
given
•
Check expire date
•
Use a new batch of Starrsed Control
E116: QC is out of acceptable range!
Troubleshooting
168
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...