When there is a sensor alarm, an alarm indicator is shown in the tab
Maintenance
.
All functions to check the status of the sensors are grouped under button
Check sensors
.
•
Check Fill stop sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
•
Check temperature sensor: Value must be equal to the actual room temperature near the pipette belt.
The value can be set in tab
Settings
.
•
Check Diluter start sensor: This sensor is only used in
. If the diluter does not start during
the aspiration, the status of this sensor must be checked.
Click the Check button, the OK sign is shown if the sensor value is in range.
•
Check Diluent flow sensor: This sensor is only used in
. When activated, the indicator
Stand-by is activated and the indicator Flow is not. When the button Test is clicked, the indicator Flow
must be shown as "activated". After finishing the test, both indicators must be shown as "activated".
•
Check Separator sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
•
Check Flow sensor: Click the Check button, the OK sign is shown if the sensor value is in range.
•
Check Measure sensor:
Clean sensor first before executing this function.
Click the Check button, the OK sign is shown if the sensor value is in range.
Press the button
Measure
. The pipette currently at the measure position will be measured.
The results are displayed in graphical form:
Measure head start
position correct
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
71
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...