6.7.1.1 - Mixer motor
The following settings need to be set;
Mixing settings;
1.
Number of turns for long mix
.
▪ This is the first mix of a new sample rack. The minimum mix value is set default for 8 mixing.
2.
Number of turns short mix
.
▪ After the Mixing interval time a second mix is started, the number of the second mix is set default
for 1 mix.
Check both mixing settings with the Test button. Observe the numbers in the numerical field.
Mixing motor settings;
1.
Mixing direction:
▪ Default setting is left.
2.
Mixing speed:
default setting 90%
3.
Hold level:
default setting 30%
4.
Homing speed:
default 20%
After setting use Save settings to save the new settings. If Save settings is not pressed the old settings
will be used
Starrsed program