1. Blocked filter. Replace blue disk filter.
2. Waste separator wrongly assembled or blocked.
▪
.
3. Main vacuum pump defective.
▪
Fatal error, only allowed
for a trained technician
.
4. Bad vacuum. Adjustment needed on the vacuum.
▪
Fatal error, only allowed
for a trained technician
.
5. Blockage in the 3 way vacuum manifold.
▪
Fatal error, only allowed
for a trained technician
.
15.6 - Flushing liquids
After each sample aspiration the entire system is washed automatically.
If there is no liquid flow:
•
Check that the peristaltic pumps are running. If the pump tubes are worn or leaking, replace the
tubes.
•
Check that the pump tubes are installed correctly.
•
Check the tubes between the containers and pumps/valves.
•
Unscrew the cap from the bottle or cubitainer. Check the pick-up tubes in the bottle or cubitainer.
•
Check the tubes for blockages or kinks.
15.6.1 - De-ionized water
Select from
Maintenance -> Prime / Clean
-> [Prime De-ionized water]
, the vacuum pump should operate and liquid
flows through the thin tube connected to the side of the fill nozzle cap.
After each sample aspiration, the fill nozzle is washed automatically with de-ionized water.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
151
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...