Work instruction Number
172
Page
1 of 2
Purpose:
Cleaning Measure sensor
Safety:
Bio Hazard area
Instrument
: Starrsed RS
Revision:
002, February 2014
Remove of the top cover
1. Switch Instrument
OFF
.
2. Remove the two rear screws of the top cover.
3. Lift the top cover carefully from the ESR unit.
If the measure sensor is out of range, the sensor must be cleaned.
(172)
In order to clean the measure sensor remove the pipette at the measuring position (complete with top and bottom
clamp).
For cleaning use a cotton bud dipped in deionised water or aerosol air blower, make sure the cotton bud is just
damp. Do not use any organic solvents.
Pipette removal
1. Push and pull vertically the pipette from the
holding position of the belts.
2. Take pipette off the carousel.
3. Store the pipette on a safe place.
Switch Starrsed RS
ON
1. Carefully clean the inner part of the measuring sensor by using a cotton bud.
2. Check the values of the Measure sensor MS
50
±10 by using the
Check measure sensor
function.
3. If not in range repeat cleaning the inner part of the measuring sensor.
4. When in range switch
OFF
the Starrsed RS.
Pipette installation
Work instruction Starrsed RS
212
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...