Work instruction Number
223
Page
1 of 1
Purpose:
Tube sensor detection and barcode reader
settings
Safety:
None Bio Hazard area
Revision:
001,December 2013
The tube detection sensor is positioned in a way that it detects light which is reflected from the plastic tube caps.
Some tube caps (for example black or green caps) may not reflect enough light to activate the tube sensor.
This can be solved by adjusting the light intensity of the tube detection sensor or to activate the switch
Always read
barcode
.
A. Adjust tube detection sensor:
1. Select
Service - Rack unit - Position motor
.
2. Switch
ON
the
Check tube detection sensor
3. Insert a sample rack with a sample tube on position "1" in the grabber.
4. Press
IN position
and put the rack inside the grabber.
5. Press
Mix position
, rack is now held in the grabber
6. Press
TUBE position
"1", mixer is now moving to the "1" position.
7. Adjust the sensor with a screw driver until the sensor LED changes from orange to green and also a “
Tube
detected” is shown next to the green block on the software screen.
8. Press
Mix position
. The sensor indicator on the software screen must not light up, “
No tube !” is shown next to
the grey block. The sensor LED is still orange.
B. Activate
Always read barcode
1. Select
Service - Rack unit - Position motor
.
2. Switch on
Always read barcode
.
3. Insert a sample rack with a sample tube on position 1 in the grabber.
4. Press
IN position
and put the rack inside the grabber.
5. Press
Mix position
, rack is now held in the grabber.
6. Press
TUBE position TUBE#
1, mixer is now moving to the 1 position.
7. Press
Read barcode
, the barcode is shown above the button, even if the tube sensor detection is not detecting
a tube.
8. Check if the information is the same as on the barcode label.
Work instruction Starrsed RS
252
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...