E25: "Measure" sensor out of
range. Check/clean this sensor !
The Measure sensor has reached a
critical level.
Continuing could result in wrong
ESR results.
•
Check and/or clean the
Measure sensor. If cleaning
fails, replace sensor
E26: "Diluent Flow" sensor out of
range. Check/clean this sensor !
The EDTA Flow sensor has reached
a critical level.
Continuing could result in filling
errors.
•
Check and/or clean the
EDTA Flow sensor. If cleaning
fails, replace sensor
E27: "Temperature" sensor out of
range. Check Settings !
The measured room temperature
has reached a critical level.
Continuing could result in wrong
ESR results.
•
Check the temperature
sensor setting.
•
Check and/or clean the
Temperature sensor. If
cleaning fails, replace sensor
E29: Result path not found.
Switched to default (D:\). Check
"Result Path" setting.
Selected result path is not valid.
Software is using the default
setting
•
Check result path setting
•
Check if network or USB
devices are used.
E30: No ACK/NACK received from
host after sending inquiry!
No response from Host within a
certain time limit after sending an
inquiry 3 times.
•
Check communication cable
between Host and Starrsed RS
computer.
•
Check serial port settings
(baud rate, etc...)
•
Check protocol settings.
•
Check Host computer.
E31: NACK received from host
after sending inquiry!
Did not receive ACK from Host
after sending inquiry 3 times.
See E30
E32: LIMS Connection timeout.
Host not found!
The ESR analyser unit could not
establish a connection with the
HOST (server) via TCP/IP.
•
Check TCP/IP settings
•
Check network cable
•
Check HOST settings
E34: No response from host after
sending 'Sample data record'!
No response from Host within a
certain time limit after 3 attempts.
See E30
E35: No response from host after
sending 'Sample flag record'!
No response from Host within a
certain time limit after 3 attempts.
See E30
E36: No ACK/NACK received after
sending 'Sample result string'!
No response from Host within a
certain time limit after 3 attempts.
See E30
Appendix for Starrsed RS
MRN-071-2-EN version 7.02 Starrsed RS
189
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...