15.8.6 - Random air bubbles in pipette
1. Check Diluent flow by priming the diluter system.
2. Insufficient sample volume.
Random air bubbles result in ESR Error 3.
15.9 - Leaking pipettes
If blood or cleaning solutions leak from a pipette, perform the following procedures and check the
performance of the system after each step to see if the problem has been solved. If the completion of the
following steps does not result in a correction of the problem contact technical support.
1. Check for specks of dirt or hairs in the pipette valves.
2. A scratched valve tube.
3. A scratched valve body.
4. Valve on top of the pipette is dirty or damaged.
5. Check pipette bottom, glass may be chipped.
6. Check the pipette valve for contamination or wear.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
157