1. Blood clots or rubber debris from the tube cap in the sample.
▪ Check the condition of the outer needle.
▪ Check the mixing time of the tube
Settings
-
General settings
-
Mixing interval
<default 90 sec>
2. Filling procedure stopped by operator.
3. Insufficient sample volume.
▪ Should be at least 1.4 ml.
4. Faulty fill nozzle or fill nozzle washer/O-ring.
▪ Check fill nozzle and washer/O-ring.
5. Incorrectly adjusted sample probe depth.
▪ Check needle depth
Settings
-
General settings
-
Sample probe depth
<default 5 mm>
6. No or poor vacuum.
▪ Check vacuum
Maintenance
-
Check sensors
-
Check flow sensor
15.14 - Carousel position error
E18 is the error code for Carousel position error. Starrsed RS was not able to position the carousel. There
was a difference found in the pipette memory position table and the actual measured position of the position
sensor.
A position error can occur if the Starrsed RS is switched OFF while the carousel is moving to the next
position or when the Rinse nozzle sticks up.
Warning
: In case of the uncertainty about the position and linked patient data, is it highly
recommended to wash all samples after a position error.
Position error solutions
1. Check in tab
Settings
(password XXXXXX)- Carousel control the Rinse position.
Troubleshooting
160
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...