1. Hook pipette assembly on to the pipette belts.
2. Make sure that pipettes are correctly fitted on to the
pipette belts.
3. Visually check if all pipette valves are at the same
height.
4. Visually check the bottom of the pipette V shape ring.
5. Incorrect fitted pipettes may cause
mechanical
damage
to the instrument.
6. Check for leakage with
Fill & Clean
.
Replace the top cover:
1. Put the cover carefully over the ESR unit.
2. Fasten the two rear screws of the top cover. (If present/if needed).
Switch
ON
the Starrsed RS.
Work instruction Starrsed RS
MRN-071-2-EN version 7.02 Starrsed RS
213