Starrsed Diluent
Grey
Deionized water
Blue
Starrsed Disinfectant
White
After each reagents change, the fluid system must be primed:
1. Select
Maintenance
->
Prime / Clean
.
2. Perform the applicable prime step to fill the relevant tubes with reagent and remove air.
13.9 - Check or replace sample probe or outer needle
A faulty or broken needle can cause a fill time-out error or a dilution error.
If necessary replace the sample probe or outer needle.
Remove the top cover from the Rack
unit.
General maintenance
MRN-071-2-EN version 7.02 Starrsed RS
139