The last QC sample was not within acceptable range or
has no result
•
Press on QC icon
▪ Press "Accept" to continue sampling without
performing a new QC, continuing could produce
incorrect results.
▪ Press "Cancel" to return. Try new QC sample
tube (normal samples will be finished)
QC result out of range!
•
Perform a new QC sample, normal samples will be
finished
•
If this error persists check/clean instrument
QC sample expired!
Use a new batch of Starrsed Control
It is not possible to link this Lab ID. Lab ID is already linked!
The "Linked QC IDs" table may only contain one link to a
particular Lab ID.
Consider changing
Automatically remove linked QC ID after
result
option to YES
Last QC result was out of range! Continuing could produce incorrect results! Do you still want to
continue?
Result of last QC sample was not within acceptable range.
•
The last QC result should be evaluated by
authorized staff to decide whether the Starrsed RS
may run patient samples depending on the the nature
of errors
•
Press "Yes" to continue sampling without
performing a new QC, press "No" to return and take
appropriate action.
General errors
Barcode is not accepted
Barcode cannot be read
Data is incorrect
Check barcode
Troubleshooting
170
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...