Rinse pump
:
ON the pipette wash peristaltic pump is switched on.
OFF the pipette wash peristaltic pump is switched off.
Note:
If vacuum pump is OFF rinse solution will spill over the Auto rack unit
All pumps OFF
:
All active pumps will be switched OFF. The waste pump remains ON for one minute.
6.7.4.2 - Needle Control
This function is only operational if the rack is in a fill position.
1.
Outer needle up
: The Outer needle is sent to the up home position.
2.
Outer needle down
: The Outer needle is sent to the down position.
3.
Sample probe up
: The Sample probe is sent to the up home position.
4.
Sample probe down
: The Sample probe is sent to the down position.
5.
Fill nozzle up
: The Fill nozzle is sent to the pipette position
6.
Fill nozzle down
: The Fill nozzle is sent to the fill nozzle home position.
7.
Rinse nozzle up
: The Rinse nozzle is sent up to the pipette.
8.
Rinse nozzle down
: The Rinse nozzle is sent to the rinse nozzle home position.
LED status display
•
LED status check for the fill- and rinse nozzle is displayed.
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
105
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...