7. Time=Actual sedimentation time in minutes.
8. T=Temperature (in degrees Centigrade).
9. Error=Error message (if the Analyser detects an error).
10.
.
9.2.1 - Report 60-Minute mode
+ REPORT E
(Not to scale)
-- StaRRsed--
Date 25/04/2019
Time:
15:28
1
2
3
4
5
6
7
8
9
10
905001
84
75 CLEAR
17
60
23
EDTA
905002
14
13 Hazy<10mm
18
60
23
EDTA
905003
22
21 Hazy<25mm
19
60
23
EDTA
905004
67
61 Hazy>25mm
20
60
23
EDTA
905005
CLEAR
21
60
23
EDTA
905006
5
5
CLEAR
22
60
23
EDTA 079
905007
24
60
23
Too many borders found
905008
25
60
23
L_err(---/ 84/
75/200)
EDTA
905002/905003/905004
Sample results with hazy aspect
905005:
Sample result with a manual aspect, where the manual aspect is shown as a number
3
in column 6 of this
data record sample.
905006:
In this sample, the dilution rate has a dilution failure of 21% and that is printed as
EDTA 079.
905007
Sample results with a text error. This sample gives Too many borders found. Result of a pipette possibly filled
with air bubbles.
905008
Sample result with a text error. This sample is given limit error L_err(---/ 84/ 75/200)
Reporting
118
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...