Work instruction Number
192
Page
1 of
2
Purpose:
Fill and Clean with adapter
Safety:
Bio Hazard area
Instrument
: Starrsed RS
Revision:
002, September 2016
.
Cleaning agent preparation Starrsed RS: Fill and clean:
This cycle takes about 90 minutes.
1. Fill the clean adapter VERA119002 with hot
de-ionized water (80 °C) till the first mark in the
adapter. (180 ml).
2. Add cleaning agent (QRR 010905) till the
second marker in the adapter. (18 ml).
3. Place the two caps on the adapter and mix
well.
4. Put the adapter in the Input-pool.
5. Select
Maintenance -
Prime/Clean,
button
Fill and
clean
.
Start Fill and clean procedure:
1. Select button OK.
2. The adapter will be transported to the needle position.
3. The needle goes down and the fill and clean process is started.
4. When all the pipettes are filled, the needle goes back to the home position and the adapter is moved to the
release position and will be transported into the End-pool.