Note:
Deionized water
Add one or two drops of Starrsed Saline to the deionized water to avoid
Reagent alarm
.
Starrsed Diluent
The solution should be discarded if it becomes turbid.
If onboard containers are used:
If the diluent does become turbid, clean the diluent bottle thoroughly with 50 ml bleach
(containing approximately 5% Na-hypochlorite) and 950 ml de-ionized water (=5% solution).
Make sure that the container is thoroughly rinsed with de-ionized water after cleaning.
If large cubitainers are used:
Replace the Diluent cubitainer.
Starrsed RS with long level sensors:
To replace the bulk reagent containers:
1. Remove level sensors and spacers from the empty container.
2. Remove empty container.
3. Place new container.
4. Remove the cubitainer screw caps and pull the necks of the bottle packs out of the cardboard box.
5. Install the level sensors and spacers.
6. Add new reagent information in the software.
WARNING:
Wrongly placed pickup tubes causes incorrect results or instrument malfunction.
Make sure to place the appropriate level sensors in the cubitainers by checking the colored
labels/colored tubes matching the color codes on the cubitainers.
The sensors and the reagents have the following color codes:
Reagent
Color code
Starrsed Rinse solution
Green
Starrsed Saline
Yellow
General maintenance
138
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...