15.1 - Peristaltic pumps
The Peristaltic pumps are located under the Waste bottle housing assembly cover.
15.1.1 - Rinse solution not primed through the system
1. Check the rinse tube condition. It may be worn or leaking or incorrectly fitted.
Check the pick-up tube in the rinse bottle or cubitainer, it may have become detached from the tube
connector in the cap. See
Peristaltic pump tube replacement
2. Check rinse solution level.
▪ If the level is insufficient, a message is displayed and the alarm sounds!
3. One of the tubes carrying the rinse solution may be blocked or kinked.
15.1.2 - Rinse solution spilling over the instrument
If rinse solution spills over the top of the pipettes, the following items must be checked:
1. Vacuum pump:
Check the vacuum pressure: Check the airflow, go to
Maintenance
tab -
Check sensors
and select
Check flow
sensor
.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
143