5. Connect new tubing to both ends of the connectors.
6. Place one end of the tube in the pump plate holder.
7. Pull the new tube over the peristaltic pump rotor.
8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed RS.
4. Fill and clean
Start Fill and clean procedure:
1. Select button OK.
2. The adapter will be transported to the needle position.
3. The needle goes down and the fill and clean process is started.
4. When all the pipettes are filled, the needle goes back to the home position and the adapter is moved to the
release position and will be transported into the End-pool.
Work instruction Starrsed RS
244
MRN-071-2-EN version 7.02 Starrsed RS