15.11.1 - Hazy reports after cleaning
If pipettes are clean and measure sensor values are correct but still a high rate of hazy reports is reported:
The measure sensor housing could be worn out. Replace the measure sensor.
15.12 - Contaminated instrument
The Starrsed RS has bacterial and micro organism's growth. Clean the instrument monthly with a strong
cleaning agent and clean, in case of recurrent hazy results, the diluent system.
See instructions Fill and Clean and Cleaning the diluent system for details.
15.13 - Fill time-out error
Normally the fill sequence takes about 3 seconds. However, if the fill sequence exceeds 10 seconds, a fill
time-out error will be generated. The ESR analyser unit aborts the fill sequence and this error message will
be shown on the display and reported.
Fill time-out error may be caused by:
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
159