Waste-tube between waste pump and waste
container blocked.
Replace the tube.
Waste pump failure.
Exchange the waste pump cassette. If the error returns,
call for service.
Electrical bridge between the waste-level
electrodes.
Clean waste separator, see
15.16 - Reagents alarm
The software checks the reagent status before starting a new rack. If a level alarm is
ON
, it will not process
the new rack. If an alarm comes
ON
during a rack, it will finish to aspirate that rack (10 samples max.).
Washing dirty pipettes always continues, as to avoid that the samples are left in the pipettes.
Do not use the reagents if expired. If expired reagent has been used accidentally, the results obtained with
these reagents may only be used, when the expiry date was not exceeded more than 30 days.
Diluent
is sensitive for bacterial growth. The solution should be discarded if it becomes turbid or infected.
When using the small onboard containers: If the diluent does become turbid, clean the diluent bottle
thoroughly with 50 ml bleach (containing approximately 5% Na-hypochlorite) and 950 ml de-ionized water
(=5% solution). Make sure that the container has been thoroughly rinsed after cleaning.
Troubleshooting
162
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...