13.2.1 - Check the sensors in service mode
Vacuum pressure check
•
Go to tab
Maintenance
->
Check sensor
. Select
Check Flow sensor
box.
Flow:
0980
± 60 Abs:
0320
±10
If the flow is not in range there might be a blockage in the vacuum flow line to the flow sensor.
Fill Stop sensor check
•
Go to tab
Maintenance
->
Check sensor
. Select
Check Fill stop sensor
box.
Fill stop sensor FS 90..
140
..165
Diluter Start sensor check
•
Go to tab
Maintenance
->
Check sensor
. Select
Diluter start sensor
box.
Diluter start sensor 400-
550
-700
Measure sensor check
•
Go to tab
Maintenance
->
Check sensor.
Select
Check measure sensor
box.
Measure sensor MS
50
±10
Temperature sensor check
•
Go to tab
Maintenance
->
Check sensor.
Select
Check Temperature sensor
box.
Temperature sensor TS [Room temperature]
Diluent flow sensor check
•
Go to tab
Maintenance
->
Check sensor.
Select
Check Diluent flow sensor
box.
Press test. When test is finished, signal Standby and Flow must shown as activated.
Separator check
•
Go to tab
Maintenance
->
Check sensor.
Select
Check Separator sensor
box.
Separator sensor <200
600
>700
13.2.2 - Cleaning waste separator
The separator is designed to separate liquid from the air and can handle a lot of blood, rinse and other used
reagents from the instrument. After a period of time the separator is getting dirty and therefore it needs to
be cleaned weekly.
Detailed instructions of this procedure can be found in the Work Instruction
.
Symptoms of a dirty separator:
General maintenance
132
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...