Remote Commands
R&S
®
FSW
996
User Manual 1173.9411.02 ─ 43
14.5.14.2
Configuring EMI Markers
The commands required to configure EMI markers are described here.
Useful commands for configuring EMI markers described elsewhere:
●
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
●
CALCulate<n>:MARKer<m>[:STATe]
CALCulate<n>:DELTamarker<m>[:STATe]
●
CALCulate<n>:DELTamarker<m>:MREFerence
●
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md>
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>
CALCulate<n>:DELTamarker<m>:LINK
●
Remote commands exclusive to configuring EMI markers
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DETector
................................996
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DETector
....................................... 996
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DETector
<Detector>
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DETector
<Detector>
This command selects the detector for a specific marker during the final measurement.
If the marker is not yet active, the command also turns the marker on.
Suffix:
<n>
.
<m>
Parameters:
<Detector>
OFF
no final measurement is performed
AVER
average detector
CAV
CISPR Average detector
CRMS
RMS Average detector
POS
maximum peak detector
QPE
quasipeak detector
*RST:
OFF
Manual operation:
See
Configuring and Performing Measurements