Remote Commands
R&S
®
FSW
1033
User Manual 1173.9411.02 ─ 43
[SENSe:]BWIDth:VIDeo:AUTO
<State>
[SENSe:]BANDwidth:VIDeo:AUTO
<State>
This command couples and decouples the video bandwidth to the resolution band-
width.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
BAND:VID:AUTO OFF
Manual operation:
See
See
See
[SENSe:]BWIDth:VIDeo:RATio
<Ratio>
[SENSe:]BANDwidth:VIDeo:RATio
<Ratio>
This command defines the coupling ratio of the video bandwidth to the resolution band-
width (RBW/VBW).
Parameters:
<Ratio>
Range:
0,001 to 1000
*RST:
1
Example:
BAND:VID:RAT 3
Sets the coupling of video bandwidth to video bandwidth =
3*resolution bandwidth
Manual operation:
See
[SENSe:]BWIDth:VIDeo:TYPE
<Mode>
[SENSe:]BANDwidth:VIDeo:TYPE
<Mode>
This command selects the position of the video filter in the signal path.
Parameters:
<Mode>
LINear
The video filter is applied in front of the logarithmic amplifier.
In linear mode, measurements with a logarithmic level scale
result in flatter falling edges compared to logarithmic mode. The
reason is the conversion of linear power values into logarithmic
level values: if you halve the linear power, the logarithmic level
decreases by 3
dB.
LOGarithmic
The video filter is applied after the logarithmic amplifier
*RST:
LOGarithmic
Example:
BAND:VID:TYPE LIN
Video filter ahead of the logarithmic amplifier
Setting Basic Measurement Parameters