Common Measurement Settings
R&S
®
FSW
494
User Manual 1173.9411.02 ─ 43
For details see
Continuous Gate
Activates or deactivates continuous gating.
This setting is only available if
If no external trigger is active yet when continuous gating is activated, external trigger 1
is automatically activated as the trigger source.
Remote command:
[SENSe:]SWEep:EGATe:CONTinuous[:STATe]
Gate Period Length
Defines the length in seconds of a single gate period in continuous gating. The length
is determined from the beginning of one gate measurement to the beginning of the
next one.
Remote command:
[SENSe:]SWEep:EGATe:CONTinuous:PLENgth
Gate Period Count
Defines the number of gate periods to be measured after a single trigger event in con-
tinuous gating.
Remote command:
[SENSe:]SWEep:EGATe:CONTinuous:PCOunt
8.6.2.4
Gate Source Settings
Access
: "Overview" > "Trigger" > "Gate Settings" > "Source" tab
By default, the gate is opened by the trigger source, which is also the gate source.
However, you can also use different sources for a general trigger and the beginnning of
the individual gates. In this case, the trigger source defines when measurement is gen-
erally possible, and the gate source determines which data is actually measured.
Triggered gated measurements are only available in applications based on frequency
sweeps (not I/Q-data based), such as the Spectrum application.
For more information see
"Triggered gated measurements"
Trigger and Gate Configuration