Remote Commands
R&S
®
FSW
1057
User Manual 1173.9411.02 ─ 43
Example:
SENS:SWE:EGAT:AUTO MAN
SENS:SWE:EGAT:SOUR EXT2
SENS:SWE:EGAT:LEV:EXT2 1V
Sets the gating trigger to a level of 1 V at trigger port 2.
Manual operation:
See
[SENSe:]SWEep:EGATe:CONTinuous:PCOunt
<Amount>
Defines the number of gate periods to be measured after a single trigger event.
Parameters:
<Amount>
integer
Range:
1 to 1023
Increment: 1
*RST:
100
Example:
SWE:EGAT:CONT:PCO 50
Manual operation:
See
[SENSe:]SWEep:EGATe:CONTinuous:PLENgth
<Time>
Defines the length in seconds of a single gate period in continuous gating. The length
is determined from the beginning of one gate measurement to the beginning of the
next one.
Parameters:
<Time>
Range:
125 ns to 30 s
*RST:
5 ms
Default unit: S
Example:
SWE:EGAT:CONT:PLEN 10
Manual operation:
See
[SENSe:]SWEep:EGATe:CONTinuous[:STATe]
<State>
Activates or deactivates continuous gating.
This setting is only available if
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Setting Basic Measurement Parameters