Troubleshooting
R&S
®
FSW
1392
User Manual 1173.9411.02 ─ 43
Error code
Error text in the case of queue poll
Error explanation
2033
Printer Not Available
This message is displayed when the selected printer is not included in the list of available
output devices. A possible cause is that the required printer driver is missing or incorrectly
installed.
2034
CPU Temperature is too high
This message is displayed when the temperature of the processor exceeds 70 °C.
Table 16-4: Power Sensor errors
Status bar message
Description
Zeroing could not be per-
formed
Zeroing could not be performed because the RF power applied is too
high.
Power sensor zero failed
16.3
Troubleshooting Remote Operation
If problems arise during measurement in remote operation, try the following methods to
solve them.
Incompleted sequential commands - blocked remote channels
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish and the remote
channel to the R&S
FSW is blocked for further commands. In this case, you must inter-
rupt processing on the remote channel in order to abort the measurement.
To regain control over a blocked remote channel
Usually, if you wait a minute for the VISA connection to detect the lost connection and
clear the control channel by itself, you can then re-establish the connection again. If
this fails, try the following:
1. Press "Local" on the front panel of the R&S
FSW to return to manual operation (if
not disabled). Then re-establish the connection.
2. Send a "Device Clear" command from the control instrument to the R&S
FSW to
clear all currently active remote channels. Depending on the used interface and
protocol, send the following commands:
●
Visa
:
viClear()
●
GPIB
:
ibclr()
●
RSIB
:
RSDLLibclr()
The remote channel currently processing the incompleted command is then ready
to receive further commands again.
Troubleshooting Remote Operation