Remote Commands
R&S
®
FSW
959
User Manual 1173.9411.02 ─ 43
[SENSe:]LIST:RANGe<ri>:LIMit:STARt
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Level>
Absolute limit at the start frequency of a SEM range.
Range:
-400 to 400
*RST:
13
Default unit: dBm
Example:
LIST:RANG2:LIM:STAR 200
Manual operation:
See
"Abs Limit Start/ Abs Limit Stop"
[SENSe:]LIST:RANGe<ri>:LIMit:STATe
<State>
This command turns the limit check for all spurious emission measurement ranges on
and off.
Suffix:
<ri>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
LIST:RANG2:LIM:STAT ON
Manual operation:
See
[SENSe:]LIST:RANGe<ri>:LIMit:STOP
<Level>
This command defines an absolute limit for a spurious emission measurement range.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Level>
Absolute limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
13
Default unit: dBm
Example:
LIST:RANG2:LIM:STOP 200
Manual operation:
See
"Abs Limit Start/ Abs Limit Stop"
Configuring and Performing Measurements