Remote Commands
R&S
®
FSW
1159
User Manual 1173.9411.02 ─ 43
CALCulate<n>:MARKer<m>:PEXCursion
<Excursion>
This command defines the peak excursion (for
all
markers in
all
windows).
The peak excursion sets the requirements for a peak to be detected during a peak
search.
The unit depends on the measurement.
Application/Result display
Unit
Spectrum
dB
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<Excursion>
The excursion is the distance to a trace maximum that must be
attained before a new maximum is recognized, or the distance to
a trace minimum that must be attained before a new minimum is
recognized
*RST:
6 dB in the Spectrum application and RF displays
Default unit: DB
Example:
CALC:MARK:PEXC 10dB
Defines peak excursion as 10 dB.
Manual operation:
See
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
<State>
This command turns marker search limits on and off for
all
markers in
all
windows.
If you perform a measurement in the time domain, this command limits the range of the
trace to be analyzed.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:X:SLIM ON
Switches on search limitation.
Manual operation:
See
"Search Limits (Left / Right)"
See
"Deactivating All Search Limits"
See
Analyzing Measurements (Basics)