Remote Commands
R&S
®
FSW
1050
User Manual 1173.9411.02 ─ 43
TRIGger[:SEQuence]:LEVel:VIDeo
................................................................................ 1053
........................................................................................ 1053
......................................................................................1053
TRIGger[:SEQuence]:TIME:RINTerval
............................................................................1055
TRIGger[:SEQuence]:DTIMe
<DropoutTime>
Defines the time the input signal must stay below the trigger level before a trigger is
detected again.
Parameters:
<DropoutTime>
Dropout time of the trigger.
Range:
0 s to 10.0 s
*RST:
0 s
Default unit: S
Manual operation:
See
TRIGger[:SEQuence]:HOLDoff[:TIME]
<Offset>
Defines the time offset between the trigger event and the start of the sweep.
A negative offset is possible for time domain measurements.
For the trigger sources "External" or "IF Power", a common input signal is used for both
trigger and gate. Therefore, changes to the gate delay will affect the trigger offset as
well.
Parameters:
<Offset>
For measurements in the frequency domain, the range is 0
s to
30
s.
For measurements in the time domain, the range is the negative
sweep time to 30
s.
*RST:
0 s
Default unit: S
Example:
TRIG:HOLD 500us
Manual operation:
See
TRIGger[:SEQuence]:IFPower:HOLDoff
<Period>
This command defines the holding time before the next trigger event.
Note that this command can be used for
any trigger source
, not just IF Power
(despite the legacy keyword).
Note:
If you perform gated measurements in combination with the IF Power trigger, the
R&S
FSW ignores the holding time for frequency sweep, FFT sweep, zero span and
I/Q data measurements.
Setting Basic Measurement Parameters