Remote Commands
R&S
®
FSW
1130
User Manual 1173.9411.02 ─ 43
Suffix:
<n>
.
1..n
Parameters:
<Mode>
VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST:
VIDeo
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]
<Detector>
Defines the trace detector to be used for trace analysis.
For details see
"Mapping Samples to sweep Points with the Trace Detector"
For EMI measurements, the trace detector is used for the initial peak search only, not
for the final test. The detector for the final test is configured using
MARKer<m>:FUNCtion:FMEasurement:DETector
If the EMI (R&S
FSW-K54) measurement option is installed and the filter type "CISPR"
is selected, additional detectors are available, even if EMI measurement is not active.
For details see
Chapter 7.13.3.2, "Detectors and Dwell Time"
Suffix:
<n>
.
<t>
Parameters:
<Detector>
APEak
Autopeak
NEGative
Negative peak
POSitive
Positive peak
QPEak
Quasipeak (CISPR filter only)
SAMPle
First value detected per trace point
Analyzing Measurements (Basics)