Remote Commands
R&S
®
FSW
1011
User Manual 1173.9411.02 ─ 43
NORMal
P5
RRC
CISPr | PULSe
CISPR (6 dB) - requires EMI (R&S
FSW-K54) option
Return value for query is always
PULS
.
MIL
MIL Std (6 dB) - requires EMI (R&S
FSW-K54) option
[SENSe:]MPOWer:RESult[:LIST]?
This command queries the results of the pulse power measurement.
This command may be used to obtain measurement results in an asynchronous way,
using the service request mechanism for synchronization to the end of the measure-
ment.
If there are no results, the command returns an error.
Return values:
<PulsePower>
List of pulse powers.
The number of values depends on the number of pulses you
have been measuring.
The unit is dBm.
Usage:
Query only
[SENSe:]MPOWer[:SEQuence]
<Frequency>, <RBW>, <MeasTime>,
<TriggerSource>, <TriggerLevel>, <TriggerOffset>, <Detector>, <NoPulses>
This command configures and initiates the pulse power measurement.
The R&S
FSW caches all measurement parameters that you can set with this com-
mand. If you use the command repeatedly, the R&S
FSW only changes those settings
that you have actually changed before initiating the measurement. Thus, measurement
times are kept as low as possible.
If you synchronize the measurement with
*OPC
, the R&S
FSW produces a service
request when all frequencies have been measured and the number of individual mea-
surements has been performed.
Note that using the command as a query initiates the measurement and returns the
results if all frequencies have been measured. For more information on querying the
results see
Parameters:
<Frequency>
Defines the pulse frequency.
Range:
0 to Fmax
Default unit: Hz
<RBW>
Defines the resolution bandwidth.
Default unit: HZ
Configuring and Performing Measurements