Measurements and Results
R&S
®
FSW
300
User Manual 1173.9411.02 ─ 43
Up to three ranges can be defined for each of the six available traces.
Comment
An optional comment can be defined for the gate range settings of each trace.
Remote command:
[SENSe:]SWEep:EGATe:TRACe<t>:COMMent
Period
Length of the period to be traced. The period is the same for all traces. If you change
the period for one trace, it is automatically changed for all traces.
Make sure the defined period is not longer than the total measurement time of the cur-
rent measurement. Keep in mind that the measurement time depends on the band-
width and the number of samples (see
rent measurement time is indicated as "Meas Time" in the channel bar.
Remote command:
[SENSe:]SWEep:EGATe:TRACe<t>:PERiod
Range <x> Use
Activates tracing of the defined range during a gated measurement.
Remote command:
[SENSe:]SWEep:EGATe:TRACe<t>[:STATe<gr>]
Range <x> Start/Stop
Defines the start and stop points of the range within the tracing period. Make sure the
value for the stopping time is smaller than the length of the period.
Statistical Measurements (APD, CCDF)